Laminar Air Flows are designed to offer the highest product protection for samples and processes by providing ISO Class 5 particle free work area. The LAFs with various air-flow patterns offer proven protection for the samples and processes. The work area is continuously supplied with positive pressure HEPA filtered air flow.
General Specifications
- Triple Stage filtration
- EU4 - Prefilters (90% down to 10 micron)
- EU7 - Intermediate Filter (95% down to 3 micron)
- EU14 - HEPA Filter - Supply / Exhaust (99.999% down to 0.3 micron)
- Statically & Dynamically balanced Motor-blower with suspension arrangement to reduce noise level
- Double skin puff filled side panels
- Horizontal internal Covings at base
- Fluorescent Lights with Milky White Diffuser
- Differential Pressure Gauges : 3 Nos
- ON/OFF Switches
- Clean Down Timer with Operation HOLD indicator
- DOP Test Port5/15 Amp Sw
- itch Socket for external equipment
Operating Conditions
- Air cleanliness : ISO Class 5 (ISO 14644-1:1999 (E))
- Air Velocities : 0.45 ± 0.05 mps
- Air Flow : Vertical - Recirculatory
- Noise Level : Less than 67 dB
- Vibration Level : Minimum
- Light intensity : > 300 lux
- POWER SUPPLY : 230V AC 1-Ø 50HZ / 415V AC 3- Ø 50HZ
Options and Accessories
A wide range of configurations to suit any containment application :
- Double skin side panels with view windows
- Stainless steel perforated grille for HEPA Filter
- Front Anti Static PVC Curtains (for warehouse applications)
- Front doors (hinged/sliding/rapid roller)
- SS Stand alone table for keeping Weighing Scale
- Sodium Vapor Lamps
- Night Mode Operation
- Velocity display with alarms (audible and visible)
- Filter blocked alarms (audible and visible)
- Explosion proof electrics for Flame - proof applications
- PLC based Control System
- Once-through airflow systems
- Safe change Bag in / Bag out arrangement for filters